Astral 200mm Aperture Calorimeter

Scientech's 200 mm aperture, volume and surface absorbing calorimeters are made for use with average laser power up to 150W or diverging beam lasers. Calibration data is stored in the detector.

These calorimeters support CW or pulsed lasers, some models with peak pulse power densities up to 45 GW/cm2.

This calorimeter also requires use of an Astral Power Meter.

 

Model 360801S 380801S 380802S 388UV5S
Detector Calibration Data stored in detector stored in detector stored in detector stored in detector
Type of Absorber Surface Volume* Volume Volume
Aperture Diameter 200mm (8") 200mm (8") 200mm (8") 200mm (8")
Minimum Beam Diameter 7.5cm 7.5cm 7.5cm 7.5cm
Spectral Response 250nm - 35µm 266nm - 1.2µm 9 -11µm 193nm - 360nm
Maximum Average Power, with full illumination of absorber 150W 150W 150W 150W
Minimum Average Power 700mW 700mW 700mW 700mW
Maximum Power Density 200W/cm²

13.5W/cm2 @1064nm

10.5W/cm2 @532nm

3.85W/cm2 @355nm

79mW/cm2 @266nm

4W/cm² 50W/cm2 @ 355nm
Maximum Peak Power Density 1MW/cm²

45GW/cm2 @1064nm

35.5GW/cm2@532nm

13.5GW/cm2@355nm

265MW/cm2 @266nm

100MW/cm²

Repetitive pulses:       101MW/cm2 @355nm

Single pulse:                3.5GW/cm2 @355nm

Maximum Single Pulse Energy 1 kJ 1 kJ 1 kJ 1 kJ

Max Energy Density, J/cm2 =

Repetitive Pulses: (pulse width in sec)

1000 x (pulse width)1/2

to max of 200J/cm2

1.85J/cm2 @1064nm

1.45J/cm2 @532nm

0.5J/cm2 @355nm

10mJ/cm2 @266nm

8J/cm2 1.1J/cm2 @ 355nm

Max Energy Density, J/cm2 =

Single Pulse: (pulse width in sec)

1000 x (pulse width)1/2

to max of 200J/cm2

3.5J/cm2 @ 1064nm

2.8J/cm2 @ 532nm

1.1J/cm2 @ 355nm

21mJ/cm2 @266nm

4J/cm2 40J/cm2 @ 355nm 
Response Time 5 sec 5 sec 5 sec 5 sec
Dimensions D x L, cm 22.86 x 25.40 22.86 x 25.40 22.86 x 25.40 22.86 x 25.40
Compatible Meters

AI310, AI310D,

AI310D-USBL

AI310, AI310D,

AI310D-USBL

AI310, AI310D,

AI310D-USBL

AI310, AI310D,

AI310D-USBL